Introduction: In the present study, a prospective systematic assessment of the clinical utility of the Ambu® aScopeTM 4 Cysto Reverse Deflection with regard to image quality, maneuverability, and navigation in an outpatient and inpatient setting was performed. Materials and Methods: A prospective multicenter study was performed for evaluation of the instrument during routine cystoscopy. We evaluated the clinical performance of the instruments using a standardized user questionnaire in different categories including image quality, treatment success, imaging of all areas of the urinary bladder, quality of navigation, flexibility of the endoscope, and satisfaction with the device. Statistical analyses were performed by SPSS using the Kruskal-Wallis and Wilcoxon-Mann-Whitney tests. A p value of p ≤ 0.05 was defined as statistically significant. Results: A total of 200 cystoscopies were performed, and the questionnaire response rate was 100%. The image quality was rated as very good in 65.5% (n = 131), good in 30.5% (n = 61), and neutral in 4% (n = 8) of cases. The criteria for poor or very poor were not mentioned. The characteristic “treatment success based on image quality” was also evaluated as very good in 49% (n = 98) and good in 50.5% (n = 101). The analysis revealed a very good or good overall impression of the examiners in all cases. Replacement of the cystoscope was not necessary during any of the examinations. However, in 3 cases, technical difficulties were documented. Further analysis of the data showed that physicians with less professional experience rated the visualization of the urinary bladder (p = 0.007) and the treatment success with regard to image quality significantly worse (p = 0.007). Conclusion: The Ambu® aScope™ 4 Cysto Reverse Deflection shows high satisfaction values among users in clinical routine. In analogy to other studies with flexible endoscopes, urologists with more professional experience show higher satisfaction values than examiners with less training in flexible endoscopy.

Cystoscopy is the most common endoscopic examination in urology. It is used for diagnostic purposes, e.g., for evaluation of the bladder mucosa in case of tumors of the bladder and urethra, as well as for interventional purposes, e.g., for forceps biopsies or for removal of inserted ureteral stents. Good endoscopic visualization and device handling are essential for the examiner.

The Ambu® aScope™ 4 Cysto Reverse Deflection is a new flexible disposable cystoscope certified for diagnostic cystoscopy in November 2020. Clinical studies show that the use of disposable endoscopes is more cost-effective compared to reusable endoscopes. Beebe et al. [1], Donato et al. [2], and Pietropaolo et al. [3] demonstrate a more favorable cost-effectiveness compared to conventional reusable cystoscopes [1, 5]. The clinical advantages of single-use endoscopy also include the immediate availability of sterile disposable instruments. When using multiple endoscopes, hygienic aspects, a possibly limited number of available devices, and the time of reprocessing between two examinations must be taken into account in hospitals and outpatient offices [6].

Currently, however, there is no comprehensive validation with regard to the satisfaction of users and examiners with the handling and the device characteristics of the Ambu® aScope™ Cysto Reverse Deflection in clinical routine. Multicenter studies with other single-use devices primarily describe their clinical possible indications, e.g., based on the success rate of the removal of inserted DJ splints [7]. Despite poor flexion due to an instrument inserted in the working channel, Scotland et al. [8] described in an observational study a satisfying image quality using a single-use cystoscope. While systematic prospective-randomized trials comparing single-use cystoscopes versus reusable cystoscopes remain methodologically challenging, user satisfaction appears as a measurable effect size to determine the product quality and clinical applicability of a novel instrument. Therefore, in the present study, a systematic assessment of the clinical usability of the Ambu® aScope™ 4 Cysto Reverse Deflection integrating image quality, maneuverability, and navigation was performed in both an outpatient and inpatient setting.

Study Population and Study Period

The evaluation was performed on a total collective of 200 patients who received a cystoscopy. Data collection took place between September 2021 and January 2022 at four centers (2 inpatient centers, 2 outpatient centers). A positive ethics vote for the observational study was granted by the Ethics Committee of the University of Tübingen under the number 387/2021B02.

Patients were included who had an indication for diagnostic cystoscopy, cystoscopic follow-up of previous urinary tract disease, biopsy of the urethral mucosa/bladder mucosa, botulinum toxin injection into the urinary bladder muscles, removal of ureteral stents, as well as for bladder stone removal in the course of a routine presentation in the clinic/practice. The Ambu® aScope™ 4 Cysto Reverse Deflection was used instead of a flexible reusable cystoscope for diagnosis/intervention. After the examination, the users were asked about image quality, navigability, and maneuverability using a standardized questionnaire. Patient-related clinical and histological variables were not included in the study.

Systematic Questionnaire

Baseline variables recorded were the investigators’ professional experience in two categories (1–10 and >10 professional years), the average number of cystoscopies performed by the investigators per week (1–2; 3–5, >5), the indication for performing cystoscopy, and any additional endoscopy accessories used. In addition, the following characteristics regarding examination quality were queried using a Likert scale (very good: 1, good: 2, neutral: 3, poor: 4, and very poor: 5): image quality, treatment success based on image quality, visualization of all areas of the urinary bladder, quality of navigation within the urinary bladder, quality of illumination, flexibility of the endoscope without endoscopy accessories, satisfaction with working channel size for biopsy, and flexibility of the endoscope with inserted endoscopy accessories. In addition, the image quality of the monitor, the menu navigation, data transfer options, and the 180° rotation possibility of the live image were evaluated. Finally, the question was asked about the overall impression regarding the image quality and functionality of the monitor as well as the overall impression regarding quality and functionality of the cystoscope (see Appendix 1: Questionnaire for users). If a change to another cystoscope was necessary during the procedure or if there were problems with the Ambu® aScope™ 4 Cysto, this was recorded.

The data were analyzed using SPSS Statistics version 28.0, and the scale levels were compared using the Kruskal-Wallis test and Wilcoxon-Mann-Whitney test. A significant difference was assumed at a value of ≤ 0.05.

Baseline Variables of the User Population

With 200 cystoscopies performed, the questionnaire response rate was 100%. Due to the anonymous questionnaire, the number of different investigators was not determined. In a total of 62.5% (n = 125) of responses, investigators reported professional experience of between 1 and 10 years, and 29.5% (n = 59) reported professional experience of more than 10 years. This information was missing from 16 questionnaires (8%). A total of 25.5% (n = 51) of respondents reported performing between 1 and 2 cystoscopies per week, 24.5% (n = 49) between 3 and 5 cystoscopies per week, and 44.5% (n = 89) more than 5 cystoscopies per week.

Indication for Cystoscopy

Table 1 summarizes the indication for cystoscopy during the observation period. Ambu® aScope™ 4 Cysto Reverse Deflection was used for the initial bladder examination in 47% (n = 94) of cases and for follow-up bladder examination in 38% (n = 76) of cases. Biopsies were performed in 3.5% (n = 7) of cases. Botox injections into the urinary bladder muscles were performed in 2.5% (n = 5) of cases. Urethral stents were removed in 4% (n = 8) of the examinations performed, and bladder stones were removed in 0.5% (n = 1) of the examinations.

Table 1.

Indication for cystoscopy

Frequency%
Initial examination of the bladder 94 47.0 
Follow-up examination of the bladder 76 38.0 
Biopsy of the bladder and/or bladder neck 3.5 
Botox injection 2.5 
Removal of ureteral stent 4.0 
Removal of bladder stone 0.5 
Other interventions 4.5 
Total 200 100.0 
Frequency%
Initial examination of the bladder 94 47.0 
Follow-up examination of the bladder 76 38.0 
Biopsy of the bladder and/or bladder neck 3.5 
Botox injection 2.5 
Removal of ureteral stent 4.0 
Removal of bladder stone 0.5 
Other interventions 4.5 
Total 200 100.0 

Results of the User Survey

A detailed overview of the results of the user survey is shown in Table 2, and a summary is given in Table 3. The image quality was rated as “very good” in 65.5% (n = 131), “good” in 30.5% (n = 61), and “neutral” in 4% (n = 8) of cases. The criteria for poor or very poor were not mentioned. The characteristic “treatment success based on image quality” was also described as “very good” in 49% (n = 98) and “good” in 50.5% (n = 101). The visualization of all areas of the urinary bladder was “very good” in 48.5% (n = 97) of cases. The bending angle of the Ambu® aScope™ 4 Cysto was described as “very good” in 123 cases (42.8%) and “good” in 68 (34.7%) of cases. With inserted endoscopy accessories, the value “very good” was achieved in 16.9% and the assessment “good” in 38 cases (64.4%). In three cases of ureteral stent removal, the categories “poor” or “very poor” were selected. The working channel size for performing a biopsy was described as “very good” in 8 cases (33.3%), “good” in 12 cases (50%), and “neutral” in 3 cases (12.5%) or poor in 4.2% of cases. Switching to another cystoscope was not necessary in any examination. In the documentation of three cases with technical difficulties, it was observed that the irrigation flow could not be controlled with inserted accessories. A second three-way stopcock became necessary for this purpose. In addition, insufficient angulation of the instrument occurred during the removal of an inserted DJ splint with inserted endoscopy accessories. Another user described an insufficient inversion for evaluation of the bladder neck. The overall impression and functionality of the monitor and the single-use cystoscope were “very good” or “good” in all cases. A comparative analysis dividing the collective according to the professional years of the users showed that the visualization of all areas of the urinary bladder and the treatment success due to the image quality were rated lower in users with less professional experience (Table 4). The stratification according to the number of endoscopies performed per week shows a similar picture here: satisfaction with the Ambu® aScope™ 4 Cysto was rated better among users with a higher number of endoscopies per week (Table 5).

Table 2.

Results from the user survey

Very goodGoodNeutralBadVery badNot specified
Image quality 131 65.5% 61 30.5% 4.0% 0.0% 0.0% 0.0% 
Treatment success due to image quality 98 49.2% 101 50.8% 0.0% 0.0% 0.0% 0.0% 
Imaging of all areas of the bladder 97 49.0% 93 47.0% 3.0% 1.0% 0.0% 0.0% 
Navigation inside the bladder 127 63.5% 61 30.5% 12 6.0% 0.0% 0.0% 0.0% 
Lighting 135 67.8% 56 28.1% 4.0% 0.0% 0.0% 0.0% 
Flexibility without endoscopy accessories 123 62.8% 68 34.7% 2.6% 0.0% 0.0% 0.0% 
Working channel size for biopsy 33.3% 12 50.0% 12.5% 4.2% 0.0% 0.0% 
Flexibility with endoscopy accessories 10 16.9% 38 64.4% 13.6% 3.4% 1.7% 0.0% 
Image quality monitor 136 68.0% 48 24.0% 15 7.5% 0.5% 0.0% 0.0% 
Menu navigation 82 41.2% 98 49.2% 19 9.5% 0.0% 0.0% 0.0% 
Data transfer option 53 61.6% 27 31.4% 5.8% 1.2% 0.0% 0.0% 
180° rotation option of the live image 31 53.4% 22 37.9% 5.2% 0.0% 0.0% 1.7% 
Total impression quality + functionality Ambu aView 2 Advance 133 67.2% 64 32.3% 0.5% 0.0% 0.0% 0.0% 
Total impression quality + functionality Ambu aScope 4 Cysto 132 66% 68 34% 0.0% 0.0% 0.0% 0.0% 
Very goodGoodNeutralBadVery badNot specified
Image quality 131 65.5% 61 30.5% 4.0% 0.0% 0.0% 0.0% 
Treatment success due to image quality 98 49.2% 101 50.8% 0.0% 0.0% 0.0% 0.0% 
Imaging of all areas of the bladder 97 49.0% 93 47.0% 3.0% 1.0% 0.0% 0.0% 
Navigation inside the bladder 127 63.5% 61 30.5% 12 6.0% 0.0% 0.0% 0.0% 
Lighting 135 67.8% 56 28.1% 4.0% 0.0% 0.0% 0.0% 
Flexibility without endoscopy accessories 123 62.8% 68 34.7% 2.6% 0.0% 0.0% 0.0% 
Working channel size for biopsy 33.3% 12 50.0% 12.5% 4.2% 0.0% 0.0% 
Flexibility with endoscopy accessories 10 16.9% 38 64.4% 13.6% 3.4% 1.7% 0.0% 
Image quality monitor 136 68.0% 48 24.0% 15 7.5% 0.5% 0.0% 0.0% 
Menu navigation 82 41.2% 98 49.2% 19 9.5% 0.0% 0.0% 0.0% 
Data transfer option 53 61.6% 27 31.4% 5.8% 1.2% 0.0% 0.0% 
180° rotation option of the live image 31 53.4% 22 37.9% 5.2% 0.0% 0.0% 1.7% 
Total impression quality + functionality Ambu aView 2 Advance 133 67.2% 64 32.3% 0.5% 0.0% 0.0% 0.0% 
Total impression quality + functionality Ambu aScope 4 Cysto 132 66% 68 34% 0.0% 0.0% 0.0% 0.0% 
Table 3.

Summary of results from the user survey

NMinimumMaximumAverage valueStandard deviation
Image quality 200 1.39 0.564 
Treatment success due to image quality 199 1.51 0.501 
Imaging of all areas of the bladder 198 1.56 0.608 
Navigation within the bladder 200 1.42 0.605 
Lighting 199 1.36 0.559 
Flexibility without endoscopy accessories 196 1.40 0.540 
Working channel size for biopsy 24 1.87 0.797 
Flexibility with endoscopy accessories 59 2.08 0.772 
Menu navigation 199 1.68 0.640 
Data transfer option 86 1.47 0.663 
180° rotation option of the live image 57 1.47 0.630 
Total impression quality + functionality Ambu aView 2 Advance 198 1.33 0.483 
Total impression quality + functionality Ambu aScope 4 Cysto 200 1.34 0.475 
NMinimumMaximumAverage valueStandard deviation
Image quality 200 1.39 0.564 
Treatment success due to image quality 199 1.51 0.501 
Imaging of all areas of the bladder 198 1.56 0.608 
Navigation within the bladder 200 1.42 0.605 
Lighting 199 1.36 0.559 
Flexibility without endoscopy accessories 196 1.40 0.540 
Working channel size for biopsy 24 1.87 0.797 
Flexibility with endoscopy accessories 59 2.08 0.772 
Menu navigation 199 1.68 0.640 
Data transfer option 86 1.47 0.663 
180° rotation option of the live image 57 1.47 0.630 
Total impression quality + functionality Ambu aView 2 Advance 198 1.33 0.483 
Total impression quality + functionality Ambu aScope 4 Cysto 200 1.34 0.475 
Table 4.

Comparison of professional years

Professional yearsNAverage rankp
Image quality 1–10 125 89.17 0.133 
>10 59 99.55 
Total 184  
Treatment success due to image quality* 1–10 125 98.20 0.007* 
>10 58 78.63 
Total 183  
Imaging of all areas of the bladder* 1–10 124 97.81 0.007* 
>10 58 78.00 
Total 182  
Navigation within the bladder 1–10 125 92.06 0.847 
>10 59 93.43 
Total 184  
Lighting 1–10 125 90.21 0.408 
>10 58 95.86 
Total 183  
Flexibility without endoscopy accessories 1–10 123 87.30 0.150 
>10 57 97.41 
Total 180  
Working channel size for biopsy 1–10 14.17 0.328 
>10 15 11.50 
Total 24  
Flexibility without endoscopy accessories 1–10 42 31.46 0.227 
>10 17 26.38 
Total 59  
Image quality monitor 1–10 125 88.94 0.104 
>10 59 100.03 
Total 184  
Total impression quality + functionality Ambu aView 2 Advance 1–10 123 89.96 0.481 
>10 59 94.71 
Total 182  
Total impression quality + functionality Ambu aScope 4 Cysto 1–10 125 89.97 0.250 
>10 59 97.86 
Total 184  
Professional yearsNAverage rankp
Image quality 1–10 125 89.17 0.133 
>10 59 99.55 
Total 184  
Treatment success due to image quality* 1–10 125 98.20 0.007* 
>10 58 78.63 
Total 183  
Imaging of all areas of the bladder* 1–10 124 97.81 0.007* 
>10 58 78.00 
Total 182  
Navigation within the bladder 1–10 125 92.06 0.847 
>10 59 93.43 
Total 184  
Lighting 1–10 125 90.21 0.408 
>10 58 95.86 
Total 183  
Flexibility without endoscopy accessories 1–10 123 87.30 0.150 
>10 57 97.41 
Total 180  
Working channel size for biopsy 1–10 14.17 0.328 
>10 15 11.50 
Total 24  
Flexibility without endoscopy accessories 1–10 42 31.46 0.227 
>10 17 26.38 
Total 59  
Image quality monitor 1–10 125 88.94 0.104 
>10 59 100.03 
Total 184  
Total impression quality + functionality Ambu aView 2 Advance 1–10 123 89.96 0.481 
>10 59 94.71 
Total 182  
Total impression quality + functionality Ambu aScope 4 Cysto 1–10 125 89.97 0.250 
>10 59 97.86 
Total 184  

Significant differences marked with *.

Table 5.

Comparison cystoscopies/week

Number of cystoscopies per weekNAverage rankp
Image quality* a. 1–2/week 51 110.39 (c) 
b. 3–5/week 49 122.33 (c) 
c. >5/week 89 71.13 (a) (b) 
Total 189  <0.001* 
Treatment success due to image quality a. 1–2/week 51 87.05 0.316 
b. 3–5/week 48 101.33 
c. >5/week 89 95.08 
Total 188  
Imaging of all areas of the bladder* a. 1–2/week 50 78.48 (a) 
b. 3–5/week 48 109.17 (b) 
c. >5/week 89 94.54  
Total 187  0.006* 
Navigation within the bladder* a. 1–2/week 51 92.26 (b) 
b. 3–5/week 49 117.56 (a) (c) 
c. >5/week 89 84.15 (b) 
Total 189  <0.001* 
Lighting* a. 1–2/week 51 108.03 (c) 
b. 3–5/week 48 121.37 (c) 
c. >5/week 89 72.25 (a) (b) 
Total 188  <0.001* 
Flexibility without endoscopy accessories* a. 1–2/week 47 117.66 (c) 
b. 3–5/week 49 120.64 (c) 
c. >5/week 89 67.90 (a) (b) 
Total 185  <0.001* 
Working channel size for biopsy a. 1–2/week 11 14.00 0.070 
b. 3–5/week 10 13.25 
c. >5/week 4.50 
Total 24  
Flexibility with endoscopy accessories a. 1–2/week 14 31.68 0.282 
b. 3–5/week 13 34.65 
c. >5/week 32 27.38 
Total 59  
Image quality monitor* a. 1–2/week 51 110.59 (c) 
b. 3–5/week 49 118.83 (c) 
c. >5/week 89 72.95 (a) (b) 
Total 189  <0.001* 
Total impression quality + functionality Ambu aView 2 Advance* a. 1–2/week 50 111.10 (c) 
b. 3–5/week 48 122.13 (c) 
c. >5/week 89 69.22 (a) (b) 
Total 187  <0.001* 
Total impression quality + functionality Ambu aScope 4 Cysto* a. 1–2/week 51 113.53 (c) 
b. 3–5/week 49 121.36 (c) 
c. >5/week 89 69.87 (a) (b) 
Total 189  <0.001* 
Number of cystoscopies per weekNAverage rankp
Image quality* a. 1–2/week 51 110.39 (c) 
b. 3–5/week 49 122.33 (c) 
c. >5/week 89 71.13 (a) (b) 
Total 189  <0.001* 
Treatment success due to image quality a. 1–2/week 51 87.05 0.316 
b. 3–5/week 48 101.33 
c. >5/week 89 95.08 
Total 188  
Imaging of all areas of the bladder* a. 1–2/week 50 78.48 (a) 
b. 3–5/week 48 109.17 (b) 
c. >5/week 89 94.54  
Total 187  0.006* 
Navigation within the bladder* a. 1–2/week 51 92.26 (b) 
b. 3–5/week 49 117.56 (a) (c) 
c. >5/week 89 84.15 (b) 
Total 189  <0.001* 
Lighting* a. 1–2/week 51 108.03 (c) 
b. 3–5/week 48 121.37 (c) 
c. >5/week 89 72.25 (a) (b) 
Total 188  <0.001* 
Flexibility without endoscopy accessories* a. 1–2/week 47 117.66 (c) 
b. 3–5/week 49 120.64 (c) 
c. >5/week 89 67.90 (a) (b) 
Total 185  <0.001* 
Working channel size for biopsy a. 1–2/week 11 14.00 0.070 
b. 3–5/week 10 13.25 
c. >5/week 4.50 
Total 24  
Flexibility with endoscopy accessories a. 1–2/week 14 31.68 0.282 
b. 3–5/week 13 34.65 
c. >5/week 32 27.38 
Total 59  
Image quality monitor* a. 1–2/week 51 110.59 (c) 
b. 3–5/week 49 118.83 (c) 
c. >5/week 89 72.95 (a) (b) 
Total 189  <0.001* 
Total impression quality + functionality Ambu aView 2 Advance* a. 1–2/week 50 111.10 (c) 
b. 3–5/week 48 122.13 (c) 
c. >5/week 89 69.22 (a) (b) 
Total 187  <0.001* 
Total impression quality + functionality Ambu aScope 4 Cysto* a. 1–2/week 51 113.53 (c) 
b. 3–5/week 49 121.36 (c) 
c. >5/week 89 69.87 (a) (b) 
Total 189  <0.001* 

Significant differences marked with *.

With the introduction of single-use endoscopes in urology, alternative systems to reusable cystoscopes are available, which are currently under scientific validation in terms of quality, cost-effectiveness, and ecology. Available studies with single-use cystoscopes (e.g., for the removal of ureteral stents) have demonstrated their clincal efficacy and cost-effectiveness compared with reusable cystoscopes [1, 3]. The aim of our study was to investigate the satisfaction of users and examiners with the handling and features of the Ambu® aScope™ 4 Cysto Reverse Deflection in everyday clinical practice. The evaluation of the results showed a very good or good overall impression of the examiners in 100% of the cases. In none of the examinations performed was it necessary to switch to an alternative system in order to complete the examination successfully. In 2022, Whelan et al. [9] showed that there was no technical or optical disadvantage between the Ambu® aScope™ 4 Cysto Reverse Deflection and the sterilizable Olympus SD scope. With the exception of the 3 mm distance, where an optical advantage of the compared digital reusable cystoscope was observed. Scotland et al. [8] studied a single-use cystoscope (NeoScope) and found that flexion, image quality, and maneuverability were equivalent to the reusable cystoscopes. Disadvantageously, reduced flexion with instruments in the working channel was described for single-use cystoscopes.

In the present study, technical difficulties were noted in 3 of the 200 examinations performed. E.g., inversion was not sufficiently possible to assess the bladder neck in one case. In this case, the examination was performed by an examiner with low professional experience (1–2 cystoscopies per week, professional experience less than 10 years).

Schout et al. [10] evaluated beginners and advanced examiners with an endoscopy simulator and found that with increasing experience, the quality of flexible cystoscopy improves and the time required to perform a qualitative examination decreases. In order to perform a sufficient examination, a threshold of 122 cystoscopies to full quality in flexible cystoscopy was defined in another study [11]. The described difficulties in handling the equipment could therefore also be due to the low professional experience of the examiner. In the same context, examiners in the present study with higher professional experience gave a significantly better overall evaluation of the disposable cystoscope than examiners with less experience.

Furthermore, a lack of flexion of the instrument with inserted grasping forceps for the removal of an inserted DJ was documented. In this case, the examiner cited that the available grasping forceps for flexible cystoscopy were too short for the disposable cystoscope. Despite the fundamentally high compatibility of single-use instruments with endoscopic aids, sufficient passage of the working channel forceps and baskets should be checked as part of the acquisition. Here, the working channel of the Ambu aScope 4 Cysto has a diameter of 6.6 Fr. and is thus comparable to available flexible urethrocystoscopes.

In an earlier study, Wong et al. [12] focused their analysis on patient satisfaction with the Ambu® aScope™ 4 Cysto Reverse Deflection. All patients had previous experience with a reusable cystoscope. The authors described that 95% of the subjects would prefer the single-use cystoscope over the reuseable cystoscope due to the shorter procedure time.

Apart from the general aspects of using single-use cystoscopes or reusable endoscopes like logistic aspects and immediate availability of endoscopes [1, 5], costs and cost-effectiveness of the procedures [13, 14], and hygienic aspects [15, 17], increasing attention has recently been given to waste and environmental issues related to sterilizable systems and single-use instruments as part of sustainability efforts and “planetary health” in healthcare. Su et al. [4] reported costs for using reusable cystoscopes to remove DJ splints in a range from $165 to $1,469 in dependence of the number of investigations performed. In comparison, the cost of single-use cystoscopy was reported between $244 and $420. The authors concluded that for facilities performing fewer than 700 cystoscopies annually with a total of 20 multiple cystoscopes available, greater cost-effectiveness can be achieved using single-use cystoscopes. Davies et al. [18] investigated the carbon footprints of single-use URS and reusable systems and surprisingly found no significant differences between the devices (4.43 vs. 4.47 kg CO2 per device/use). However, the type of re-sterilization and the associated energy consumption for the reusable devices appear most relevant here. By contrast, the disposal of single-use devices is primarily responsible for the respective CO2 emissions. It should be kept in mind that apart from reprocessing and disposal of the medical products, transportation and logistic efforts represent highly relevant parameters for CO2 emissions in urological endoscopy.

Despite its prospective design, the present study is not without limitations. A main limitation of the present approach to user quality assessment is the lack of a control group, so that no comparative results with reusable endoscopes could be obtained. Noteworthy, a comparative study of reusable and single-use instruments is challenging in the context of a multicenter study given a diverse distribution of reusable endoscopes from different manufacturers, different imaging technologies (chip on tip or classic), and investigator experience in the respective centers.

While the analysis of single-use versus reusable endoscopy in urology is an open scientific field with ongoing research and discussion, the clinical data here demonstrate the clinical feasibility of single-use endoscopy using the Ambu® aScope™ 4 Cysto Reverse Deflection based on an extensive user survey.

The Ambu® aScope™ 4 Cysto Reverse Deflection shows high satisfaction values among users in clinical routine. In analogy to examinations with other flexible devices, examiners with more professional experience show higher satisfaction values than colleagues with less experience. A change to an alternative system was not necessary during the study.

We would like to thank Jörg Hennenlotter and Elisabeth Schenk for their logistic support and the facilitation of data transfer.

All procedures performed in studies involving human participants were in accordance with the ethical standards of the Institutional and/or National Research Committee and with the 1964 Helsinki Declaration and its later amendments. The study was approved and consented by the Local Ethical Committee of the University of Tübingen, Germany (project number: 387/2021B02). Written informed consent was not required in accordance with local guidelines.

S.R.: consultant/speaker honoraria for Advanced Accelerator Applications Germany GmbH, MDxHealth, Astellas Pharma, Pfizer Oncology, Bayer Oncology, Merck Pharma, Desitin, Ambu A/S, BMS Pharma, Eisai Pharma, and MSD Sharp & Dohme GmbH; A.N.: consultant/speaker honoraria for MSD, Janssen, AstraZeneca, Ispen, Amgen, Roche, GlaxoSmithKline, and Boston Scientific. T.K. and J.S.: consultant/speaker honoraria for LISA Laser, OptiMed, Ambu A/S, Boston Scientific, Olympus, Meditate, Dr. Schwabe, and Omega Pharma. T.L. has no conflicts of interest to declare.

This study was financially supported by Ambu GmbH, D-61231 Bad Nauheim. Germany.

Thomas Lütfrenk (T.L.) and Steffen Rausch (S.R.): project development, data collection, data analysis, and manuscript writing/editing. Andreas Neisius (A.N.), Johannes Salem (J.S.), and Timur H. Kuru (T.H.K.): data collection and manuscript writing/editing. All authors critically revised the manuscript and approved the final version. They agree to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.

All data generated or analyzed during this study are included in this article. Further inquiries can be directed to the corresponding author.

1.
Beebe
SC
,
Jenkins
LC
,
Posid
T
,
Knudsen
BE
,
Sourial
MW
.
Single-use grasper integrated flexible cystoscope for stent removal: a micro-costing analysis-based comparison
.
J Endourol
.
2020
;
34
(
8
):
816
20
.
2.
Donato
P
,
Honore
M
,
Zana
T
,
Pokorny
M
,
Bowes
W
,
Meyer
JP
.
Prospective trial of single-use, flexible cystoscope for ureteric double-J stent removal: cost and utility analysis
.
J Clin Urol
.
2020
;
13
(
2
):
160
3
.
3.
Pietropaolo
A
,
Hughes
T
,
Tear
L
,
Somani
BK
.
Comparison of ureteric stent removal procedures using reusable and single-use flexible cystoscopes following ureteroscopy and lasertripsy: a micro cost analysis
.
Cent Eur J Urol
.
2020
;
73
(
3
):
342
8
.
4.
Su
ZT
,
Huang
MM
,
Matlaga
BR
,
Hutfless
S
,
Koo
K
.
A micro-costing analysis of outpatient flexible cystoscopy: implications for adoption of single-use flexible cystoscopes
.
World J Urol
.
2021
;
39
(
11
):
4275
81
.
5.
Oderda
M
,
Antolini
J
,
Falcone
M
,
Lacquaniti
S
,
Fasolis
G
.
Cost-effectiveness analysis of a single-use digital flexible cystoscope for double J removal
.
Urologia
.
2020
;
87
(
1
):
29
34
.
6.
Ofstead
CL
,
Heymann
OL
,
Quick
MR
,
Johnson
EA
,
Eiland
JE
,
Wetzler
HP
.
The effectiveness of sterilization for flexible ureteroscopes: a real-world study
.
Am J Infect Control
.
2017
;
45
(
8
):
888
95
.
7.
Doizi
S
,
Kamphuis
G
,
Giusti
G
,
Palmero
JL
,
Patterson
JM
,
Proietti
S
.
First clinical evaluation of a new single-use flexible cystoscope dedicated to double-J stent removal (Isiris™): a European prospective multicenter study
.
World J Urol
.
2017
;
35
(
8
):
1269
75
.
8.
Scotland
K
,
Wong
VKF
,
Chan
JYH
,
Tawfiek
E
,
Chiura
A
,
Chew
BH
.
Evaluation of a single-use flexible cystoscope: a multi-institutional international study
.
J Endourol
.
2020
;
34
(
9
):
981
6
.
9.
Whelan
P
,
Kim
C
,
Tabib
C
,
Preminger
GM
,
Lipkin
ME
.
Evolution of single-use urologic endoscopy: benchtop and initial clinical assessment of a new single-use flexible cystoscope
.
J Endourol
.
2022
;
36
(
1
):
13
21
.
10.
Schout
BMA
,
Muijtjens
AMM
,
Hendrikx
AJM
,
Ananias
HJK
,
Dolmans
VEMG
,
Scherpbier
AJJA
.
Acquisition of flexible cystoscopy skills on a virtual reality simulator by experts and novices
.
BJU Int
.
2010
;
105
(
2
):
234
9
.
11.
MacKenzie
KR
,
Aning
J
.
Defining competency in flexible cystoscopy: a novel approach using cumulative sum analysis
.
BMC Urol
.
2016
;
16
(
1
):
31
7
.
12.
Wong
A
,
Phan
Y
,
Thursby
H
,
Mahmalji
W
.
The first UK experience with single-use disposable flexible cystoscopes: an in-depth cost analysis, service delivery and patient satisfaction rate with Ambu® ascope™ 4 cysto
.
J Endolum Endourol
.
2021
;
4
(
1
):
e29
44
.
13.
Herr
HW
.
The risk of urinary tract infection after flexible cystoscopy in patients with bladder tumor who did not receive prophylactic antibiotics
.
J Urol
.
2015
;
193
(
2
):
548
51
.
14.
Herr
HW
.
Should antibiotics be given prior to outpatient cystoscopy? A plea to urologists to practice antibiotic stewardship
.
Eur Urol
.
2014
;
65
(
4
):
839
42
.
15.
Noronha
AM
,
Brozak
S
.
A 21st century nosocomial issue with endoscopes
.
BMJ
.
2014
348
g2047
.
16.
Wendelboe
AM
,
Baumbach
J
,
Blossom
DB
,
Frank
P
,
Srinivasan
A
,
Sewell
CM
.
Outbreak of cystoscopy related infections with Pseudomonas aeruginosa: New Mexico, 2007
.
J Urol
.
2008
;
180
(
2
):
588
92
; discussion 592.
17.
Chang
CL
,
Su
LH
,
Lu
CM
,
Tai
FT
,
Huang
YC
,
Chang
KK
.
Outbreak of ertapenem-resistant enterobacter cloacae urinary tract infections due to a contaminated ureteroscope
.
J Hosp Infect
.
2013
;
85
(
2
):
118
24
.
18.
Davis
NF
,
McGrath
S
,
Quinlan
M
,
Jack
G
,
Lawrentschuk
N
,
Bolton
DM
.
Carbon footprint in flexible ureteroscopy: a comparative study on the environmental impact of reusable and single-use ureteroscopes
.
J Endourol
.
2018
;
32
(
3
):
214
7
.

Additional information

Johannes Salem and Timur H. Kuru: shared last authorship.